Laserfiche WebLink
a. If any of the samples in the preparation batch are non -detect (i.e., below the MDL <br />specified in the planning document), the spiking level must not be greater than 2 times <br />the Contract specified PQL. <br />b. The concentration of a spiked sample cannot exceed 5 times the highest concentration <br />of any contracted sample in the preparation batch. <br />5. Evaluation of Matrix Spikes - The results of matrix spikes must meet the acceptance criteria <br />specified by the Contract and listed in the planning document or the data must be <br />appropriately qualified. <br />a If the failure is reported to be due to sample matrix interference, the laboratory shall <br />document the process by which this conclusion is determined. <br />6. Evaluation of Laboratory Duplicate/Replicate Samples — All replicate samples (sample <br />duplicates, matrix spike duplicates, LCS duplicates or other replicates) must be evaluated <br />for a precision criterion not to exceed 20 % RPD. This criterion shall be listed in the <br />planning document <br />a In the event that laboratory replicate agreement is not observed, the laboratory must <br />investigate the poor precision and report the results with appropriate qualifiers and/or <br />comments <br />7. Instrument Calibration — In addition to calibration procedures specified in the analytical <br />methods listed in the planning document the CONTRACTOR shall ensure that the following <br />requirements are met: <br />a. All sample results shall be chronologically bracketed between acceptable calibration <br />verifications. <br />b. Initial Calibration Requirements <br />(1) The minimum number of calibration standards required to calibrate each <br />instrument used for the contracted analyses shall conform to the analytical method <br />approved in the planning document. If the minimum number of calibration standards <br />is not specified in the method, the number must be specified in the planning <br />document and shall be consistent with the NELAC Chapter 5 standards. <br />(ii) Unless otherwise specified by the method, all sample results shall be based on <br />the initial calibration curve responses. <br />(iii) If linear regressions are used, the correlation coefficient shall be equal to or <br />greater than 0.995 for all regressions. <br />(iv) Immediately after performing an initial calibration, the accuracy of the calibration <br />shall be verified using a second source. A second source may be a standard, a <br />Standard Reference Material (SRM), or other sample type with a verified <br />concentration such as a QC Check Sample. Standards must have been prepared <br />from a different lot or vendor. <br />(v) The acceptance criteria for second -source verifications shall be specified in the <br />planning document. <br />(vi) Sample analysis cannot proceed if an initial calibration is unacceptable. <br />c. Continuing Calibration Requirements: <br />(i) When an initial calibration is not performed on the day of analysis, a continuing <br />calibration standard shall be analyzed, evaluated and determined to be acceptable <br />prior to analyzing samples. <br />(ii) A continuing calibration standard shall be analyzed and evaluated at the end of <br />the analytical run. <br />(iii) The acceptance criteria for continuing calibration verifications shall be specified <br />in the planning document <br />(iv) For each analytical run, the analytical sensitivity must be evaluated using a <br />continuing calibration standard prepared at the Contract -specified PQL. The <br />analyzed value of this standard must be within 70% — 130% of the expected value. If <br />DEP Agreement No. LP###, Attachment H, Page 10 of 12 <br />